Nano Archive

An integrated microfluidic chip for non-immunological determination of urinary albumin

Lin, Chun-Che and Tseng, Chin-Chung and Huang, Chao-June and Wang, Jung-Hao and Lee, Gwo-Bin (2010) An integrated microfluidic chip for non-immunological determination of urinary albumin. BIOMEDICAL MICRODEVICES, 12 (5). pp. 887-896. ISSN 1387-2176 (Print) 1572-8781 (Online)

Full text is not hosted in this archive but may be available via the Official URL, or by requesting a copy from the corresponding author.

Official URL: http://www.springerlink.com/content/n5m7154656v4r2...

Abstract

This study presents an integrated microfluidic chip for non-immunologically determining the concentrations of albumin in clinical urine samples. This microchip integrates membrane-type micromixers and a fan-shaped micropump capable of simultaneously and precisely delivering assay reagents to react with 6 urine samples in one single operation. The experimental results show that the coefficient of variation in the pumping rate is 2.42%. More importantly, using this unique chip design, only 2 electromagnetic valves are required for the actuation of the micromixer and the micropump. The working range of the proposed microchip is 2–200 mg/L of albumin, which covers the range of interest for the determination of microalbuminuria. Moreover, statistical analysis show that the results obtained by the proposed microchip are in good agreement with the conventional detection method, based on immunological assays. This simple, inexpensive and microchip-based platform presents a promising alternative to conventional immunological assays for measurement of urinary albumin, and is well suited for clinical applications.

Item Type:Article
Subjects:Biomedical Science > Nanobiotechnology
ID Code:9477
Deposited By:M T V
Deposited On:10 Aug 2010 15:57
Last Modified:16 Aug 2010 15:52