Graham, Anthony H. D. and Bowen, Chris R. and Taylor, John and Robbins, Jon (2009) Neuronal cell biocompatibility and adhesion to modified CMOS electrodes. Biomedical Microdevices, 11 (5). pp. 1091-1101.
| PDF 447Kb |
Abstract
The use of CMOS (Complementary Metal Oxide Semiconductor) integrated circuits to create electrodes for biosensors, implants and drug-discovery has several potential advantages over passive multi-electrode arrays (MEAs). However, unmodified aluminium CMOS electrodes may corrode in a physiological environment. We have investigated a low-cost electrode design based on the modification of CMOS metallisation to produce a nanoporous alumina electrode as an interface to mammalian neuronal cells and corrosion inhibitor. Using NG108-15 mouse neuroblastoma x rat glioma hybrid cells, results show that porous alumina is biocompatible and that the inter-pore distance (pore pitch) of the alumina has no effect on cell vitality. To establish whether porous alumina and a cell membrane can produce a tight junction required for good electrical coupling between electrode and cell, we devised a novel cell detachment centrifugation assay to assess the long-term adhesion of cells. Results show that porous alumina substrates produced with a large pore pitch of 206Â nm present a significantly improved surface compared to the unmodified aluminium control and that small pore-pitches of 17Â nm and 69Â nm present a less favourable surface for cell adhesion.
| Item Type: | Article |
|---|---|
| Subjects: | Biomedical Science > Nanobiotechnology |
| ID Code: | 8003 |
| Deposited By: | CSMNT |
| Deposited On: | 17 Aug 2010 14:30 |
| Last Modified: | 17 Aug 2010 14:30 |
Repository Staff Only: item control page

