Turcu, Florin and Schulte, Albert and Schuhmann, Wolfgang (2004) Scanning electrochemical microscopy (SECM) of nanolitre droplets using an integrated working/reference electrode assembly. Analytical and Bioanalytical Chemistry, 380 (5-6). pp. 736-741. ISSN 1618-2642
| PDF - Published Version 324Kb |
Official URL: http://www.springerlink.com/content/yrd24ah28j29ra...
Abstract
Scanning electrochemical microscopy (SECM) has been performed in the restricted space of nanolitre droplets with a robust and easy-to-handle coaxial electrode assembly centring a Pt microdisk in a circular Ag electrode. Straightforward and reproducible fabrication of the specially designed probe tips was achieved by using Tollens' reaction to chemically deposit a uniform and well-adhering layer of silver on the body of a glass-insulated Pt microdisk electrode. The suitability of the novel dual-electrode SECM tip for measurement in small volumes was evaluated by imaging an array of four Pt band microelectrodes in 500 nL electrolyte.
| Item Type: | Article |
|---|---|
| Subjects: | Analytical Science > Microscopy and probe methods Analytical Science > Metrology and standards in nanotechnology |
| ID Code: | 1523 |
| Deposited By: | SPI |